Correct positioning of central venous catheters using a new electric method

Correct positioning of central venous catheters using a new electric method

Post author correction

Article Type: ORIGINAL ARTICLE

Article Subject: Dialysis

DOI:10.5301/jva.5000361

Authors

Maurits K. Konings, Leonard J. van Schelven, Ronald C. Meijer, Albert H. Westra, Roland A. Snijder, Anton J. van Boxtel

Corresponding author

  • Maurits K. Konings
  • Department Of Medical Technology and Clinical Physics
  • Room Azu - C 01.230
  • University Medical Center Utrecht
  • 3584 Cx Utrecht, The Netherlands
  • m.konings@umcutrecht.nl

Abstract

In order to find the correct final position of the tip of a central venous catheter, we have developed a new electric method (the Proximity of Cardiac Motion (PCM) method), designed to work in tandem with the existing ECG-based method.

A small, patient-safe, high-frequency current is fed through the catheter (via the saline-filled lumen of the catheter, or a stylet). Simultaneously, the resulting voltage is measured by two electrodes on the frontal thoracic skin. The catheter tip hence functions as a current source inside the vasculature. The cardiac motion produces a variation in the amplitude of the measured voltage in the rhythm of the cardiac cycle, and the strength of this oscillatory variation is proportional to the strength of the incident current field on the heart, which is a rapidly decaying function of the distance between the catheter tip and the cavoatrial junction (CAJ). Hence the strength of this oscillatory variation is a strong indicator for the proximity of the catheter tip with respect to the CAJ.

The new method has been tested in an animal model, yielding an average final position of the catheter tip of 2.1 cm above the CAJ, with a maximum deviation of 0.5 cm.

We conclude that the new PCM method can be combined with the existing ECG method, and may potentially have significant added value when the ECG method cannot be applied, for example, in patients with atrial fibrillation or a pacemaker.

Article History

Disclosures

Financial support: This research has not received financial support.
Conflict of interest: No conflict of interest.

This article is available as full text PDF.

  • If you are a Subscriber, please log in now.

  • Article price: Eur 36,00
  • You will be granted access to the article for 72 hours and you will be able to download any format (PDF or ePUB). The article will be available in your login area under "My PayPerView". You will need to register a new account (unless you already own an account with this journal), and you will be guided through our online shop. Online purchases are paid by Credit Card through PayPal.
  • If you are not a Subscriber you may:
  • Subscribe to this journal
  • Unlimited access to all our archives, 24 hour a day, every day of the week.

Authors

  • Konings, Maurits K. [PubMed] [Google Scholar] 1, * Corresponding Author (m.konings@umcutrecht.nl)
  • Schelven, Leonard J. van [PubMed] [Google Scholar] 1
  • Meijer, Ronald C. [PubMed] [Google Scholar] 2
  • Westra, Albert H. [PubMed] [Google Scholar] 1
  • Snijder, Roland A. [PubMed] [Google Scholar] 1
  • Boxtel, Anton J. van [PubMed] [Google Scholar] 3

Affiliations

  • Department of Medical Technology, University Medical Center Utrecht, Utrecht - The Netherlands
  • Department of Thoracic Surgery, University Medical Center Utrecht, Utrecht - The Netherlands
  • Infu-In, Bilthoven - The Netherlands

Article usage statistics

The blue line displays unique views in the time frame indicated.
The yellow line displays unique downloads.
Views and downloads are counted only once per session.

No supplementary material is available for this article.